T201 Gas Filled Thermometers Remout Mounting

Wide temperature range (from -200°C up to 700°C)

Protect personnel and process from accidental contamination

Nontoxic

Suitable in the food, pharmaceutical, chemical and petrochemical industries, cryogenics, etc.

CASE DIAMETER

100 and 150mm

ACCURACY CLASS

1 According EN 13190 at 20°C

SCALE RANGES

From 200 °C to +700°C

OVER RANGE

Short time 20% full scale up to 500°C

10% full scales for ranges >500°C

STEM PRESSURE RATING

25 bar max (without thermowell)

 

STANDARD FEATURES

CASE AND RING

Material : AISI 304 Stainless Steel

Style : bayonet ring

Protection class : IP55(EN 60529)

TEMPERATURE ELEMENT

Inert gas pressure system (nontoxic) consisting of : bourdon tube,

Helical type, bulb in AISI316, seamless capillary in AISI316

POINTER

Adjustable Black aluminum

MOVEMENT

AISI304 stainless steel

DIAL

Aluminum with black markings on white background

WINDOW

Glass4mm

BULB

Bulb size: see next page table Sensitive portion and its extension with same rigid diameter

OPERATING TEMPERATURE

Middle 2/3 of the dial scales

CONNECTIONS

Sliding on stem

Sealing rings for working temperature :

PTFE (Teflon) from 100°C to + 250°C

Copper up to 700°C

THREAD SIZES

1/2 “ BSP or NPT– M

UNIT OF MEASUREMENT

Degree Celsius or Fahrenheit

AMBIENT TEMPERATURE AT CASE

0…40°C max

Others on request

AMBIENT ERROR

Ambient error is a function of line length and ambient temperature (0,25% at middle scale of span per

10°C change in ambient temperature)

 Bi-metallic link to minimize errors due to ambient temperature variations

LINELENGTH

100 ft(30 mt) max

CAPILLARY MATERIAL

Stainless steel AISI316 with stainless steel flexible armor Ø7 mm

TEMPERATURE LIMITS FOR STORAGE AND TRANSPORT

50…+70°C without liquid damping

20…+60°C with liquid damping

TEMPERATURE RANGES

See table ranges

OPTIONAL EXTRAS

Case with liquid damping max 250°C

Glycerin 99,7%

Silicon oil

Fluorolube

Case with food compatible liquid damping

Case and ring material AISI316

External adjustment ±5 % F.S.

Back and flange material AISI316

Stem diameter 6,8,10 or 14mm

Protection class: IP65 EN 60529

Safety glass window

Dual scales °C /°F

Other ranges available on request

Other thread sizes

ATEX version II 2GD c

Electrical contacts magnetic type

Inductive contact for use in hazardous areas

PTB99ATEX2219X/

PTB00ATEX2049B

ZELM63ATEX0128X

THERMOWELLS

Protection wells are recommended for applications where the stem may be exposed to pressure, corrosive or high velocity fluids. The use of thermowell also permits instrument interchange or calibration check without disturbing or shutting down the process stream

BULB TYPE AND DIMENSIONS

The bulb is manufactured in AISI 316 seamless pipe and it’s made in two parts : sensitive portion (length depending on bulb diameter) and rigid extension (same diameter of sensitive part).

Bulb can also be in electro welded AISI 316 pipe.

MINIMUM LENGTH OF SENSING PORTION (see below table)

Valid for direct mounting or with capillary  10and>10m

                           ≤ 10m                 >10m

 

 

 

 

 

 

 

 

CAPILLARY TYPES

Capillary AISI 316 Ø2,5mm - 260 °C / + 700°C

Capillary AISI 316 covered with AISI 316 armour Ø7mm - 260 °C /+400°C

Capillary AISI 316 covered with AISI 316 armour PVC coated 60°C /+120°C

Print | Sitemap
© manotherm inc